Dual source: AC adapter (Universal 240V ~ 50/60Hz input) + removable lithium battery pack. BMS + MCU power-detect switches sources instantly — no restart, all functions identical on battery. Both sources lost → system off; power back → boots into the saved configuration (power loss never resets settings).
Setup & Operation Guide
1. Flip the POWER switch ON. The WCS logo boots the system → UNCONFIGURED / WAITING SETUP.
2. Admin access (2 steps): hold UP (+) + DOWN (−) together for 9 seconds → screen shows ADMIN UNLOCK READY → press SILENT 3 times within 10 seconds (dots ●●● fill on screen).
3. Assign the cabinet mode + voice: in Admin Mode, scroll the cabinet list with UP / DOWN, press START/STOP to confirm → then choose the voice language (ENGLISH / ARABIC / OFF) with UP / DOWN and press START again. The cabinet boots with your selection. (Press SILENT to cancel; admin auto-exits after 30s idle.)
4. Door alarm (all cabinets): any door open → screen shows a live countdown → buzzer + ALARM: DOOR OPEN! at 60 seconds. Alarm stops when doors close.
5. Restocking: while a door is open, press MAX LIGHT (light 100%) or LIGHT OFF (light off) → door alarm and light pulsing are suppressed. Closing the door returns everything to normal; reopening restarts the 60s countdown.
6. Silent modes: short press SILENT → SILENT (AUTO) with 60-min countdown, then auto back to audible. Hold SILENT 5s → SILENT (LOCK) — stays silent until you hold 5s again (back to AUTO). Critical alarms (gas / overheat) always sound, even in silent mode. Silent also stops the interior-light pulsing (the exterior strip keeps pulsing). Demo time: 1s ≈ 1min.
7. Voice announcements: the speaker announces every operational event — door open/closed, door-release guidance (102), cleaning start/finish, temperature & gas alarms, maintenance, silent-mode changes. The voice language is set in Admin: after confirming the cabinet type, the screen shows ADMIN: VOICE — cycle ENGLISH / ARABIC / OFF with UP/DOWN and press START. The choice is saved and kept on every power-on. Silent modes mute the voice (same rule as the buzzer); critical alarms always speak. Admin mode itself is silent. The Live Component Monitor has a Speaker row showing the exact text being announced.
8. Sound test: press 🔊 SIM: Sound test (below) — you should hear 3 soft beeps, then the voice. The screen shows the audio state: AUDIO: RUNNING = OK. If it shows SUSPENDED, the browser is blocking sound — in Safari: Settings for This Website → Auto-Play → Allow All Auto-Play, then reload.
9. Dual power (all cabinets): every unit runs on the AC wall adapter (Universal 240V ~ 50/60Hz input) and/or a removable lithium battery pack — the BMS + MCU power-detect switches between them instantly, with no restart, and every function works the same on either source. If BOTH sources are lost (AC unplugged + battery removed or empty), the system powers off completely; when power returns it boots and continues with the saved configuration — power loss never resets settings (the only full reset is re-assigning the cabinet in Admin). Test it: click AC ADAPTER or BATTERY PACK in the Power Supply strip below the console.
Maintenance & Service (user-level — no admin needed): • Yellow MAINT LED + screen message = maintenance required. After fixing the issue, hold MAX LIGHT + LIGHT OFF together for 5 seconds to reset. Safety rule: the reset only clears the record — if the physical condition still exists (e.g. temp still ≥50°C), the alarm re-triggers immediately. • HEPA filter life (cabinets 102 / 106): 30 days. Demo time is compressed: 1 real second ≈ 1 simulated hour (~12 min to expire).
Cabinet Mode List (assign in Admin: scroll UP/DOWN → START/STOP):
101 - X Cabinet102 - PassBox Cabinet103 - Endoscope Storage & Drying104 - Safety Cabinet105 - Instruments Cabinet106 - Instruments Cabinet Pro107 - Hot and Cold Trolley Set108 - Medical Blanket Warmer